Product Image
More sustainable than new
£220.00£437.00Save 50%

Product Details

Manufacturer

About this product

ReadyCircuit single-use sensor assemblies, SciLog pressure sensor - Press Sens 0.75IN TCRM 1PK 

Single-use tubing assemblies integrated with sensors for temperature, pressure and conductivity and compatible with SciLog BioProcessing Systems SciTemp™, SciPres™, and SciCon™ monitors (part numbers 080-790, 080-690, and 080-590, respectively).
  • Sensors and ReadyMate single-use connectors are preconfigured into sensor assemblies for immediate use in a ReadyCircuit assembly.
  • Sensors assemblies are supplied gamma-irradiated with a certificate of quality.
  • Accurate and reliable pressure, temperature, and conductivity sensors with easy connection to read-out monitors.
  • Sensor data can be logged via the sensor monitor interfaces.

Sensor assemblies

ReadyCircuit Sensors are preconfigured assemblies of standard temperature, pressure and conductivity sensors, ReadyMate single-use connectors, and where applicable, reinforced silicone tubing.

Fluid path sizing

Some of the standard components include 3/8 inch pressure, temperature, and conductivity sensors with silicone tubing, and 3/8 inch ReadyMate connectors. The larger ¾ inch diameter sensors utilize direct fractional T/C ReadyMate connections.

Monitoring systems

All sensors are compatible with their relevant "parent" SciLog, SciTemp, SciPres, or SciCon monitors.

Sensor connector and cables are lockable and waterproof.


Product specificationsSensor Assembly, SciLog pressure sensor

Animal Free Compliance Animal-free Origin for wetted parts per EMEA/410/01 (alternatively-manufacturing under denaturing conditions). 
Autoclavable Yes 
Certificate of Analysis Yes 
Compatibility Compatible with most sanitizing agents such as NaOH, hypochlorite, for flow cell sanitization. 
Connections ReadyMate 
Endotoxin For process fluid contact surfaces, endotoxin level is <0.25 EU/ml using water extraction method. 
Hemolysis Testing Hemolysis, ISO 10993-4 
Material (Fluid Contact) Medical grade Polysulfone (PS) meets USP 88 Class VI and FDA 21CFR177.1520. 
Material Compliance USP Biological Safety, Class VI for wetted parts, USP 88 plastics requirements and FDA 21CFR177.1520. 
Sensor Isolation Insoluble silicone dielectric gel isolates sensing element from process solution. The gel is a non toxic, nonallergenic elastomer system. 
Sensor Microchip EPROM-Stored sensor ID and calibration factor. 
Sensor Type Medical grade, silicone piezoresistive sensing element with on-chip temperature compensation 
Sterility Each component is sterile for all fluid-contact surfaces (per AAMI TIR 33:2005 and the principles of ISO/AAMI/ASTM 11137-1:2006). 
Temperature Range 0°C to 60°C 
Tubing i.d. 1.9 mm (0.75 in)

 

Catalog Number: 12410233

Expiration Date: 21/01/2023

Related Products

What Our Clients Say

Supporting innovation at leading research companies

RNAnalytics

RNAnalytics

Leading The Way In Gene Therapy Analytics

As a startup focused on developing analytical toolkits for LNP characterization, RNAnalytics operates under tight budget constraints while striving to maintain a fully stocked lab with high-quality materials. Sustainability is also a core pillar of our operations, pushing us to make environmentally conscious decisions wherever possible....